4-42
Assembly Replacement
A8 Data Generator Board (Options UN3 & UN4)
A8 Data Generator Board (Options UN3 & UN4)
Tools Required
T-10 TORX screwdriver
long nose pliers
Removal
1. Remove the instrument cover. (Refer to "Instrument Cover" in this chapter.)
2. Remove the top cover by removing the 11 screws that secure it.
3. Disconnect A17W1,W3, W4, W5 , and W19 from the data generator board (A8).
4. Pull up on the data generator board to remove it from the motherboard connector.
Replacement
1. Reverse the removal procedure and do the following:
Refer to Chapter 3, "Replaceable Parts," to verify that the cables are reconn
ected in the correct locations.
Torque all T-10 TORX screws to 9 in-lbs.