4-24
Assembly Replacement
A2 Display
A2 Display
Tools Required
T-10 TORX screwdriver
5/16-inch open-end wrench
Removal
1. Remove the front panel assembly. (Refer to "Front Panel" in this chapter.)
2. Remove the small bracket (item 1) that secures W9 (not shown) to the RFI shield (item 2). Disconnect W9.
3. Remove the RFI shield by removing the seven screws (item 3) that secure it.
4. Remove the display (A2) by removing the four screws (item 4) that secure it.
Replacement
1. Reverse the removal procedure and do the following:
Torque all T-10 TORX screws to 9 in-lbs.
Torque W9 to 9 in-lbs.