Agilent InfinityLab LC Series VWD User Manual 116
8Test Functions
Dark Current Test
Figure 35 Dark Current Test - Signals
Dark Current Test Failed
Dark Current Test Failed
Probable cause Suggested actions
1
Defective sample or reference diode.
Replace optical unit.
2
Defective sample or reference ADC board.
Replace optical unit.
3
Defective main board.
Exchange the main board.