Agilent InfinityLab LC Series VWD User Manual 94
7Error Information
Detector Error Messages
Grating or Filter Motor Errors
Error ID: Grating: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809; Filter:
7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
During the motor tests, the detector moves the motor to the end position while
monitoring the end-position sensor. If the end position is not found, the error
message is generated.
Wavelength test failed
Error ID: 7890
The automatic wavelength check after lamp ignition has failed. When the lamp is
switched on, the detector waits 1 min to warm-up the lamp. Then a check of the
deuterium emission line (656 nm) via the reference diode is performed. If the
emission line is more than 3 nm away from 656 nm, the error message is
generated.
Test 0 Failed: Filter motor.
Test 1 Failed: Grating motor.
Probable cause Suggested actions
1
Motor is not connected.
Please contact your Agilent service representa-
tive.
2
Defective motor.
Please contact your Agilent service representa-
tive.
3
Defective/missing grating or filter.
Please contact your Agilent service representa-
tive.
4
Cable/connector defective.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Calibration incorrect.
Recalibrate the detector.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7114B and is the answer not in the manual?

Agilent Technologies G7114B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7114B
CategorySecurity Sensors
LanguageEnglish

Related product manuals