Agilent InfinityLab LC Series VWD User Manual 119
8Test Functions
Holmium Oxide Test
Holmium Oxide Test Failed
Holmium Oxide Test Failed
Probable cause Suggested actions
1
Detector not calibrated.
Recalibrate the detector.
2
Dirty or defective flow cell.
Repeat the test with the flow cell removed. If the
test is OK, exchange the flow cell components.
3
Dirty or defective holmium oxide filter.
Run the holmium oxide filter test. If the test fails,
contact your Agilent service representative.
4
Optical misalignment.
Please contact your Agilent service representa-
tive.