Agilent InfinityLab LC Series VWD User Manual 117
8Test Functions
Holmium Oxide Test
Holmium Oxide Test
This test verifies the calibration of the detector against the three wavelength
maxima of the built-in holmium oxide filter. The test displays the difference
between the expected and measured maxima. The figure below shows a
holmium test spectrum.
The Holmium Oxide Test is available in Agilent Lab Advisor (preferred tool).
The test uses the following holmium maxima:
360.8 nm
418.5 nm
536.4 nm
When to do the Test
after recalibration,
as part of the Operational Qualification/Performance Verification procedure,
or
after flow cell maintenance or repair.
NOTE
See also “Declaration of Conformity for HOX2 Filter” on page 220.