Agilent InfinityLab LC Series VWD User Manual 88
7Error Information
General Error Messages
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam
is removed, the fan is switched off, and the error message is generated.
ERI Messages
Error ID: 11120 (+5 V) , 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to over
current situations on the +5 V and +24 V lines.
Probable cause Suggested actions
1
The top foam was removed during operation.
Please contact your Agilent service representa-
tive.
2
Foam not activating the sensor.
Please contact your Agilent service representa-
tive.
3
Defective sensor or main board.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
The load on the ERI is too high.
Reduce the load.