Agilent InfinityLab LC Series VWD User Manual 145
9Maintenance
Replace Leak Handling System Parts
Replace Leak Handling System Parts
Parts required p/n Description
5043-0856 Leak Adapter
5063-6527 Tubing assembly, i.d. 6 mm, o.d. 9 mm, 1.2 m (to waste)
Preparations
Open or remove the doors
1 Locate the Leak Adapter [1] and Tubing [2]
Next Steps:
2 Press the Leak Adapter [1] down and remove it together
with the tubing.
3 Install the Leak Adapter by pressing it into the Main
Cover.
4 Insert the Tubing [2] (ca. 85 mm required for replace-
ment) between Leak Adapter outlet and Leak Pan.
5 Insert/close the doors.