Agilent InfinityLab LC Series VWD User Manual 135
9Maintenance
Replace the Flow Cell / Cuvette Holder
Replace the Flow Cell / Cuvette Holder
For flow cell details see
“Standard Flow Cell 10 mm / 14 µl” on page 150
“Micro Flow Cell 3 mm / 2 µl” on page 152
“Semi-micro Flow Cell (Parts)” on page 154
“High Pressure Flow Cell (Parts)” on page 156
When
If application needs a different type of flow cell or the flow cell needs repair.
Tools required p/n Description
Wrench, 1/4 inch
for capillary connections
OR 5043-0915 Fitting mounting tool
Parts required # Description
1 Flow cell
Preparations
Turn the lamp OFF.
1 Open the doors. 2 Disconnect the inlet capillary and outlet tubing from the
flow cell.