Replace the Deuterium Lamp
7 Insert the lamp (RFID tag on top) (1.) and fix the screws
(2.).
8 Reconnect the connector.
9 Replace the fan cover (1.) and fix its screw (2.). 10 Close the doors.
Next Steps:
11 Reset the lamp counter as described in the User Interface documentation (required for non-RFID tag lamps only).
12 Turn the lamp ON and give the lamp more than 10 minutes to warm-up.
13 Perform a Wavelength Re-calibration after lamp warm-up.
After lamp on, the detector requires a warm-up time of 60 min. No
measurements should be performed during this time.