EasyManuals Logo
Home>Agilent Technologies>Security Sensors>G7114B

Agilent Technologies G7114B User Manual

Agilent Technologies G7114B
222 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Agilent InfinityLab LC Series VWD User Manual 23
1 Introduction to the Variable Wavelength Detector
Leak and Waste Handling
Leak Sensor
Waste Concept
1 Agilent recommends using the 6 L waste can with 1 Stay Safe cap GL45 with
4 ports (5043-1221) for optimal and safe waste disposal. If you decide to use
your own waste solution, make sure that the tubes don't immerse in the liquid.
CAUTION
Solvent incompatibility
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor.
The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible
with DMF.
Do not use DMF.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7114B and is the answer not in the manual?

Agilent Technologies G7114B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7114B
CategorySecurity Sensors
LanguageEnglish

Related product manuals