Agilent InfinityLab LC Series VWD User Manual 97
7Error Information
Detector Error Messages
Heater at fan assembly failed
Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is switched on or
off a heater self-test is performed. If the test fails an error event is created. As a
result the temperature control is switched off.
Heater Power At Limit
Error ID: 1074
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has been
hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause Suggested actions
1
Defective connector or cable.
Please contact your Agilent service representa-
tive.
2
Defective heater.
Please contact your Agilent service representa-
tive.
Probable cause Suggested actions
1
Excessive ambient temperature change.
Wait until temperature control equilibrates.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7114B and is the answer not in the manual?

Agilent Technologies G7114B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7114B
CategorySecurity Sensors
LanguageEnglish

Related product manuals