Agilent InfinityLab LC Series VWD User Manual 143
9Maintenance
Correcting Leaks
Correcting Leaks
Figure 42 Correcting leaks
Leak sensor area [1]
Flow cell area [2]
When
If a leakage has occurred in the flow cell area or at the capillary connections.
Tools required Description
Tissue
Wrench, 1/4 inch
for capillary connections
1 Open the doors. 2 Use tissue to dry the leak sensor area [1].
3 Observe the capillary connections and the flow cell area
[2] for leaks and correct, if required.