Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 105
Optimizing the Detector
5
Method Development
Table 12 Conditions for simultanoeus multi wavelength detection for PNA-analysis (see
figure below)
Column Vydac, 2.1 x 250 mm, PNA, 5 µm
Mobile phase A = water; B = acetonitrile (50 : 50)
Gradient 3 min, 60 %
14.5 min, 90 %
22.5 min, 95 %
Flow rate 0.4 mL/min
Column temperature 22 °C
Injection volume 2 µL
FLD settings PMT 12,
response time 4 s
The upper trace was
received with
traditional
wavelength
switching.
Figure 36 Simultaneous multi wavelength detection for PNA-analysis
%HQ]DDQWKUDFHQH
&KU\VHQH
%HQ]REIOXRUDQWKHQH
%HQ]RNIOXRUDQWKHQH
%HQ]DS\UHQH
'LEHQ]RDKDQWKUDFHQH
%HQ]RJKLSHU\OHQH
,QGHQRFGS\UHQH
1DSKWKDOHQH
$FHQDSKWKHQH
)OXRUHQH
3KHQDQWKUHQH
$QWKUDFHQH
)OXRUDQWKHQH
3\UHQH
H[FLWDWLRQ:/DWQP
HPLVVLRQ:/DW
DQGQP
([ (P 77
5HIHUHQFH
FKURPDWRJUDP
ZLWKVZLWFKLQJHYHQWV
7LPH>PLQ@