Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 209
Maintenance
9
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector that
can be carried out without opening the main cover.
Table 28 Simple repairs
Procedure Typical Frequency Notes
Flow cell exchange If application requires a different flow cell type or if
defective.
Complete Assembly
A wavelength calibration check should
be performed after replacement.
If the flow cell is removed and inserted,
then a quick calibration check is
performed.
If this fails, you must do a wavelength
recalibration, see “Wavelength
Verification and Calibration” on
page 182.
Flow cell flushing If flow cell is contaminated.
Leak sensor drying If leak has occurred. Check for leaks.
Leak handling System
replacement
If broken or corroded. Check for leaks.