Error Information
7
General Error Messages
Cover Violation
Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
ERI Messages
Error ID: 11120 (+5 V) , 11121 (+25 V)
The ERI (Enhanced Remote Interface) provides two error events related to
over current situations on the +5 V and +24 V lines.
Probable cause Suggested actions
1
The top foam was removed during
operation.
Please contact your Agilent service
representative.
2
Foam not activating the sensor.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
The load on the ERI is too high.
Reduce the load.

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals