EasyManuals Logo

Agilent Technologies G7121A User Manual

Agilent Technologies G7121A
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #178 background imageLoading...
Page #178 background image
178 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Wavelength Accuracy Test
Wavelength Accuracy Test
Using the Agilent Lab Advisor
1 Set up the HPLC system and the Agilent Lab Advisor.
2 Flush the flow cell with clean bi-distilled water.
3 Turn on the FLD lamp.
4 Run the Wavelength Accuracy Test.
5 The FLD will change into the multi-excitation mode with emission
wavelength at 397 nm and scan in the range of the expected maximum of
350 nm±20 nm.
As result, the maxima should be found at 350 nm ±3nm, see Figure 62 on
page 178.
The FLD will change into the multi-emission mode with excitation
wavelength at 350 nm and scan in the range of the expected maximum of
397 nm±20 nm.
As result, the maxima should be found at 397 nm ±3nm, see Figure 62 on
page 178.
Figure 62 Excitation and Emission Spectrum (expected results)
(; QP
IL[HGIL[HG
(0 QP
(; QP
QP
(0 QP
QP

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals