EasyManuals Logo

Agilent Technologies G7121A User Manual

Agilent Technologies G7121A
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 123
Optimizing the Detector
5
Finding the Best Wavelengths
A Real Example
Although an excitation wavelength of 340 nm is quoted in the literature the
module scan of orthophthalaldehyde, a derivative of the amino acid alanine,
(Figure 45 on page 123) shows a maximum between 220 nm and 240 nm.
Figure 45 Scan Orthophthalaldehyde Derivative of Alanine
When you are looking for the wavelength by scanning, scan over the whole
range. As this example shows a maximum may be found in a completely
different wavelength range.
NOTE
When comparing fluorescence excitation spectra directly with DAD spectra or literature
based absorbance spectra, you should consider large differences in the used optical
bandwidth (FLD = 20 nm) which cause a systematic wavelength maximum shift depending
on the absorbance spectrum of the compound under evaluation.

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals