Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 213
Maintenance
9
Exchanging a Flow Cell
Exchanging a Flow Cell
For bio-inert modules use bio-inert parts only!
When If an application needs a different type of flow cell or the flow cell is defective (leaky).
Tools required p/n Description
Wrench, 1/4 inch
for capillary connections
OR 5043-0915 Fitting mounting tool
Parts required p/n Description
G1321-60005 Flow cell, 8 µL, 20 bar (pH 1 – 9.5)
G1321-60015 Flow cell, 4 µL, 20 bar (pH 1 – 9.5)
G5615-60005 Bio-inert flow cell, 8 µL, 20 bar (pH 1–12) includes Capillary Kit Flow Cells BIO
(p/n G5615-68755)
G1321-60007 FLD Cuvette Kit, 8 µL, 20 bar
Preparations Turn off the flow.
CAUTION
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
Do not mix bio-inert and non-inert modules or parts in a bio-inert system.

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals