214 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
9
Maintenance
Exchanging a Flow Cell
New instruments are shipped with a flow cell installed and have a factory wavelength
calibration.
If a flow cell is used other then the installed/shipped flow cell:
• verify that the installed flow cell is fixed very tight,
• perform a wavelength calibration using Glycogen (part of FLD Calibration Kit
(G7121-68001)).
Before storing the flow cell, flush it with iso-propanol and close it with Plastic fittings
(0100-1259).
1
Open the doors of the module.
2
Disconnect the capillaries from the flow cell.