88 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
5
Optimizing the Detector
Getting Started and Checkout
Getting Started and Checkout
This chapter describes the check out of the Agilent 1260 Infinity II
Fluorescence Detector using the Agilent isocratic checkout sample.
Starting Your Detector
1 Turn ON the detector.
2 Turn ON the lamp.
When the lamp is turned on the first time the instrument performs some
internal checks and a calibration check which takes about 5 minutes.
3 You are now ready to change the settings of your detector.
Some features (for example, spectrum acquisition, multi-wavelength detection) described
in this chapter are not available on the G1321C FLD.
When If you want to checkout the detector
Parts required # p/n Description
1 5063-6528 Start-up Kit, includes
1 LC Column and parts listed below
1 01080-68704 Agilent isocratic checkout sample
This 0.5 mL ampoule contains 0.15 wt.% dimethylphthalate, 0.15 wt.%
diethylphthalate, 0.01 wt.% biphenyl, 0.03 wt.% o-terphenyl in
methanol.
1 0100-1516 Fitting male PEEK, 2/pk
1 5021-1817 Capillary ST 0.17 mm x 150 mm
Hardware required LC system with FLD