118 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
5
Optimizing the Detector
How to collect spectra with modes SPECTRA ALL IN PEAK and APEX SPECTRA ONLY
Notes:
The peak-detection algorithm works best when a peak is reduced to 8 – 16
data points. The FLD collects the data points with an internal data rate of
74.08 Hz (= 13.50 ms) (1 signal only). The data reduction is only influenced
by the PDPWparameter.
The peak-detector works online on the current chromatogram. This means
that begin/apex/end of a peak is recognized with delay. Additionally the
points of spectra are sequentially acquired. This means that the acquisition
of wide range spectra lasts much longer than the acquisition of a
short-range spectrum. When you have a fast chromatography it is nearly
impossible to collect a "clean" APEX-spectra: the first/last points of the
spectra are acquired before/after you have the highest concentration in the
detector’s cell.
How long the acquisition of single spectra lasts is shown in the FLD’s setup
screen.

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals