Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 129
Optimizing the Detector
5
Finding the Best Signal Amplification
As a result, the "ADC overflow" is visible as a real flat peak in the
chromatogram showing the user, that the setting of the detector parameter
(PMT gain or the concentration of the solution) is set to high.
NOTE
The transfer of methods 1:1 from one FLD to another may result into the above "ADC
overflow" problem. For details see “FLD Scaling Range and Operating Conditions” on
page 125.