188 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
8
Test Functions and Calibration
Wavelength Calibration Procedure
2 Preparation of the Flow Cell.
a Flush the flow cell with water.
b Remove the inlet capillary from the flow cell.
c Take the syringe and fix the needle to the syringe adapter.
d Suck about 1.0 ml of the calibration sample into the syringe.
e Keep the syringe in a horizontal position.
f Remove the needle.
g Add the filter to the syringe and fit the needle to filter.
Figure 67 Syringe with Sample Filter
h Lift the needle tip and carefully eject approximately 0.5 ml to remove air
out of the syringe and to flush the needle.
i Add the PEEK fitting to the needle tip and fix both at the flow cell inlet.
j Slowly inject about 0.2 ml and wait for about 10 seconds to inject another
0.1 ml. This will assure that the cell is filled properly.
3 Wavelength Calibration.
a From the user interface start the FLD wavelength calibration.
Agilent Lab Advisor: Calibrations
Instant Pilot (G4208A): Maintenance > FLD > Calibration
6DPSOHILOWHU
NOTE
Do not inject the calibration sample without the sample filter.
NOTE
If the wavelength calibration process fails, refer to “Wavelength Calibration Failed” on
page 159.

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals