220 Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
9
Maintenance
Correcting Leaks
1
Open the doors.
2
Use tissue to dry the leak sensor area [1].
3
Observe the capillary connections and the flow cell area
[2] for leaks and correct, if required.
4
Close the doors.