Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 67
Using the Fluorescence Detector
3
Advanced Settings
Analog Output
The range can be set to either 100 mV100 or 1 V1 full scale,
see Detector Control Settings.
Zero Offset: 1 – 99 % in steps of 1 % (5 % equal to
50 mV5).
Attenuation: 0.195 – 1600 LU at discrete values for
either 100 mV100 or 1 V full scale.
Multiple Wavelengths (G7121B ONLY)
You can choose to use three additional wavelengths, either
excitation or emission wavelengths in your analysis.
Choose Off to switch off multiple wavelengths, or Multi Ex
or Multi Em to set up multiple wavelengths.
See also Signal A.