Agilent 1260 Infinity II Preparative Autosampler User Manual 155
Maintenance
9
Exchange the Needle Assembly
1
In the Agilent Lab Advisor software select Service &
Diagnostics in the system screen (Tools) Maintenance
Positions > Change Needle/Loop, click Start and wait
until the needle assembly is in maintenance position.
2
Open the front door.
3
Unscrew the safety cover screw.
Sharp needle
Uncovered needles may cause injuries
➔
Do not touch the tip of the needle.
4
Remove the safety cover.