Chapter 4: Testing Performance
Offset Accuracy Test
44
6
Repeat step 5 for the remaining vertical sensitivities for channel 1 in the Zero Error
Test section of the Performance Test Record.
7 Press Default Setup, then turn off channel 1 and turn channel 2 display on.
8 Configure the scope to measure V avg on Channel 2 as follows:
a Pull down the Utilities menu and select Acquisition. When the Acquisition Setup window
is displayed, enable averaging and set the # of averages to 256.
b Change the vertical sensitivity of channel 2 to 5 mV/div.
c Click the V avg measurement icon on the left side of the screen.
d When the Enter Measurement Info window is displayed, ensure that the Vavg function
is set up as follows and then click OK:
Source = Channel 2
Measurement area = Entire Display
9
Press the Clear Display key on the scope, wait for the #Avgs value to return to 256 and
then record the scope’s mean V avg reading in the Zero Error Test section of the
Performance Test Record.
10 Repeat step 9 for the remaining vertical sensitivities for channel 2 in the Zero Error
section of the Performance Test Record.
11 Repeat steps 7 through 10 for channels 3 and 4.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infiniium 80000 Series and is the answer not in the manual?

Agilent Technologies Infiniium 80000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniium 80000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals