Chapter 4: Testing Performance
Performance Test Record
62
Offset Gain Test
Vertical
Sensitivity
V
Aux Out
Setting
V
DMM+
V
DMM-
V
DMM0
V
Scope+
V
Scope-
V
Scope0
Calculated
Offset Gain
Error
Offset Gain
Error Test
Limits
Channel 1
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 2
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 3
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %
Channel 4
5 mV/div ±400 mV ±2 %
10 mV/div ±400 mV ±2 %
20 mV/div ±400 mV ±2 %
50 mV/div ±900 mV ±2 %
100 mV/div ±1.6 V ±2 %
200 mV/div ±2.4 V ±2 %
500 mV/div ±2.4 V ±2 %
1 V/div ±2.4 V ±2 %

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Infiniium 80000 Series and is the answer not in the manual?

Agilent Technologies Infiniium 80000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniium 80000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals