Chapter 5: Troubleshooting
Primary Trouble Isolation
75
4 When you are finished, click Close.
If any of the LEDS do not work, go to “To check the LEDs” later in this chapter.
5 If both tests pass, go to step F.
F
Self Calibration
1 Complete a self Calibration by following the procedures in chapter 3, “Testing
Performance.”
2 If the calibration test fails, replace the acquisition assembly. If the calibration test passes,
go to step G.
G
The system is operational. Performance test the oscilloscope using the procedures in
chapter 3 of this service manual.