General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 83
Agilent 8485D Prior to the power meter being zeroed and calibrated, an Agilent
11708A 30 dB reference attenuator and an APC 3.5 (f) to 50 Ω (m)
N-Type adapter (08485-60005) should be connected between the
power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Agilent 8487A This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) to connect to the power meter. Remove this attenuator
before making measurements.
Agilent 8487D Prior to the power meter being zeroed and calibrated, an Agilent
11708A 30 dB reference attenuator and an APC 2.4 (f) to 50 Ω (m)
N-Type adapter (08487-60001) should be connected between the
power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
Sensor Model Connection Requirements

Table of Contents

Other manuals for Agilent Technologies N1914a

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1914a and is the answer not in the manual?

Agilent Technologies N1914a Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1914a
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals