EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>N1914a

Agilent Technologies N1914a User Manual

Agilent Technologies N1914a
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
Using E9300 E-Series Power Sensors 3
N1913/1914A EPM Series Power Meters User’s Guide User’s Guide 105
Power Meter Configuration
The N1913/1914A EPM Series power meters automatically recognize an
E- Series E9300 power sensor when it is connected. The sensor calibration
data is automatically read by the power meter. The power meter also
configures the auto- averaging settings shown in Figure 3- 1 to suit the
power sensor characteristics.
Figure 3-1 E9300 E-Series auto-averaging settings
-4 dBm
-10 dBm
Sensor Power
Minimum
Sensor Dynamic Range
10 dBm
2 dBm
1234
1114
Resolution Setting
Number of Averages
11416
11
8
32
1416128
11664128
Maximum
-40 dBm
-50 dBm
-20 dBm
-30 dBm
1114
11216
121664
4 16 128 256
32 64 256 256
Upper Range
Lower Range
Sensor Power
6 dBm
0 dBm
20 dBm
12 dBm
-30 dBm
-40 dBm
-10 dBm
-20 dBm
26 dBm
20 dBm
40 dBm
32 dBm
-10 dBm
-20 dBm
10 dBm
0 dBm
E9300/1B
E9300/1H
E9300/1/4A
NOTE
These values are valid only for the power meter channel connected with E-Series E9300
power sensor. You can also configure the settings manually – refer to Achieving Stable
Results with TDMA Signals” on page 113 if required.

Table of Contents

Other manuals for Agilent Technologies N1914a

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1914a and is the answer not in the manual?

Agilent Technologies N1914a Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1914a
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals