6 Using N8480 Series Power Sensors
164 N1913/1914A EPM Series Power Meters User’s Guide
2
The 8482B and 8482H power sensors use the same data as the 8482A.
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9. These tables do not contain any data when the power meter is
shipped from the factory.
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First, select the table you want to edit as follows:
1 Press , , to display the Sensor Tbls
screen.
Figure 6-11 “Sensor Tbls” screen
2 Choose the table you want to edit using the and keys. Press
to display the Edit Cal screen as shown in Figure 6- 12.