10 Specifications and Characteristics
218 Agilent N1913/1914A EPM Series Power Meters User’s Guide
Power Meter Specifications
Frequency Range
9 kHz to 110 GHz, power sensor dependant
Power Range
–70 dBm to +44 dBm (100 pW to 24 W), power sensor dependant
Power Sensors Compatibility
• Agilent 8480 Series power sensors
• Agilent E9300 E–Series average power sensors
• Agilent E4410 E–Series average power sensors
• Agilent N8480 Series power sensors
• Agilent U2000 Series average USB power sensors
Single Sensor Dynamic Range
• 90 dB maximum (Agilent E4410 E- Series power sensors)
• 80 dB maximum (Agilent E9300 E- Series power sensors)
• 50 dB maximum (Agilent 8480 Series power sensors)
• 55 dBm maximum (Agilent N8480 Series power sensors)
• 44 dBm maximum (Agilent U2000 Series USB power sensors)
Display Units
Absolute: Watts (W) or dBm
Relative: Percent (%) or dB
Display Resolution
Selectable resolution of: 1.0, 0.1, 0.01 and 0.001 dB in logarithmic mode,
or 1, 2, 3, and 4 significant digits in linear mode
Default Resolution
0.01 dB in logarithmic mode or three digits in linear mode

Table of Contents

Other manuals for Agilent Technologies N1914a

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1914a and is the answer not in the manual?

Agilent Technologies N1914a Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1914a
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals