EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>N1914a

Agilent Technologies N1914a User Manual

Agilent Technologies N1914a
263 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
3 Using E9300 E-Series Power Sensors
108 N1913/1914A EPM Series Power Meters User’s Guide
Procedure
1 Zero and calibrate the power meter/sensor combination.
2 Ensure the power sensor is disconnected from any signal source.
3 Press and the channel softkey. The Zeroing pop- up is
displayed.
4 Connect the power sensor to the POWER REF output.
5 Press and the channel softkey to start the calibration
routine. The Calibrating pop- up is then displayed.
Now set the frequency of the signal you want to measure. The power
meter automatically selects the appropriate calibration factor.
6 Press . On dual channel meters select the required channel.
7 Use the and keys to highlight the Frequency value field and
press
to display the Frequency pop- up. Use the numeric keypad to
enter the required value in the Frequency pop- up window.
Zero
Cal
NOTE
You can reduce the steps required to carry out the zero and calibration procedure as
follows:
1 Connect the power sensor to the POWER REF output.
2 Press and . (For dual channel meters, press
, or as required).
Zero + Cal
Zero + Cal
Zero + Cal A
Zero + Cal B

Table of Contents

Other manuals for Agilent Technologies N1914a

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1914a and is the answer not in the manual?

Agilent Technologies N1914a Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1914a
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals