Maintenance 9
Agilent N1913A/1914A EPM Series Power Meters User’s Guide 201
–
213
Init ignored
Indicates that a request for a measurement initiation was ignored as the power meter
was already initiated.
For example, INIT:CONT ON
INIT.
–214
Trigger deadlock
TRIG:SOUR was set to HOLD or BUS and a READ? or MEASure? was attempted,
expecting TRIG:SOUR to be set to IMMediate.
–220
Parameter error;Frequency list must be in ascending order.
Indicates that the frequencies entered using the MEMory:TABLe:FREQuency command
are not in ascending order.
–221
Settings conflict
This message occurs under a variety of conflicting conditions. The following list gives a
few examples of where this error may occur:
If the READ? parameters do not match the current settings.
If you are in fast mode and attempting to switch on for example, averaging, duty cycle or
limits.
Trying to clear a sensor calibration table when none is selected.
–222
Data out of range
A numeric parameter value is outside the valid range for the command.
For example, SENS:FREQ 2 kHZ.
–224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command. You
may have used an invalid parameter choice.
For example, TRIG:SOUR EXT.
–226
Lists not same length
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set to ON and the
frequency and calibration/offset lists do not correspond in length.