Maintenance 9
Agilent N1913A/1914A EPM Series Power Meters User’s Guide 205
–
310
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In this instance the
power meter detects power sensors connected to both it’s front and rear channel inputs.
–351 Configuration memory lost; storage fault
Refer to “Memory Erase/Secure Erase” on page 91
to perform secure erase.
–321
Out of memory
The power meter required more memory than was available to run an internal operation.
–330
Self-test Failed;
The -330, “Self-test Failed” errors indicate that you have a problem with your power
meter. Refer to “Contacting Agilent Technologies” on page 208 for details of what to do
with your faulty power meter.
–330
Self-test Failed;Measurement Channel Fault
–330
Self-test Failed;Measurement Channel A Fault
–330
Self-test Failed;Measurement Channel B Fault
–330
Self-test Failed;Calibrator Fault
Refer to “Calibrator” on page 194 if you require a description of the calibrator test.
–330
Self-test Failed;ROM Check Failed
–330
Self-test Failed;RAM Check Failed
–350
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
–361
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity
cannot be guaranteed.