Troubleshooting
Troubleshooting Assembly Level Problems
1-39
If the voltages are not present, replace the A9 Interface Board.
If the voltages are present, proceed to step 9.
If A9J2-4 is not 12V, measure A9J2-3.
If J2-3 is 12V, replace the A9 Interface Board.
If J2-3 is not 12V, go to step 9.
Figure 1-14 A9 Interface Board (A9J2-3 and A9J2-4 Test Points)
9. Refer to Figure 1-15. Using the voltmeter measure A3J6-4.
If A3J6-4 is 12V replace the W1 ribbon cable.
If A3J6-4 is not 12V go to step 10.
A9J2-3
A9J2-4
A9J2-