Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Troubleshooting Performance Test Problems
1-74
FM Distortion Replace the A3 RF assembly
Replace the A3 RF assembly
a
.
Digital Modulation Quality
(N5162A/82A only)
Replace the A2 Vector module N/A
DCFM Carrier Offset Replace the A3 RF assembly
Replace the A3 RF assembly
a
.
Amplitude Switching Speed Replace the A3 RF assembly
Replace the A7 Micro Deck
a
.
If this does not correct the problem, replace the
A3 RF assembly
a
.
Frequency Switching Speed Replace the A3 RF assembly
Replace the A3 RF assembly
a
.
a
After replacing the A7 Micro Deck or the A3 RF assembly, go to Chapter 4, "Post-Repair Procedures and Performance Verification".
Performance Test that is Failing... Action to Perform...
(N5161A/81A/62A/82A)
Action to Perform...
(N5183A)