2-4 Service Guide N5230-90014
General Product Information PNA Series Microwave Network Analyzers
Analyzer Options Available N5230A
For high dynamic range measurements, front panel jumpers are moved to reverse the
signal path through one of the couplers, allowing for a 15 dB improvement in transmitted
signal sensitivity in one direction only. Jumpers are installed on both ports allowing the
user to choose a measurement in either the forward or reverse direction.
Option 080, Frequency Offset Mode
This upgrade enables the additional firmware necessary to make frequency offset
measurements.
Option 1CM, Rack Mount Flange Kit for Instruments without Handles
Option 1CM provides a rack mount kit that contains a pair of flanges, rack mount rails,
and the necessary hardware to mount the analyzer in an equipment rack with 482.6 mm
(19 inches) horizontal spacing.
Additional kits may be ordered from Agilent. Refer to “Rack Mount Kits and Handle Kits”
on page 6-35 for the part numbers.
Option 1CP, Rack Mount Flange Kit for Instruments with Handles
Option 1CP provides a rack mount kit that contains a pair of flanges (cut to adapt to
handles), rack mount rails, and the necessary hardware to mount the analyzer in an
equipment rack with 482.6 mm (19 inches) horizontal spacing.
Additional kits may be ordered from Agilent. Refer to “Rack Mount Kits and Handle Kits”
on page 6-35 for the part numbers.
Option 1E1, Source Attenuators
This option is included in the active measurement configuration.
This option adds a 60-dB step attenuator in the signal path of each of the measurement
ports. These step attenuators are used to adjust the power level (in 10 db steps) to the
device under test (DUT) without changing the power in the reference path.
Option 221, 2-Port Front Panel Upgrade
This option changes the front panel UI (user interface) to a touchscreen and softkey
interface with a new color LCD display. This option changes your analyzer from an “A”
model (N5230A) to a “C” model (N5230C).
Option 897, Built-In Tests for Commercial Calibration
This option enables performance tests that are built into the analyzer that enable you to
perform ISO 9001 commercial calibration of your analyzer. Refer to “Calibration Options”
on page 2-10 for a description of the commercial calibration.