8-10 Service Guide N5230-90014
General Purpose Maintenance Procedures PNA Series Microwave Network Analyzers
Error Terms N5230A
Error Term Data
The error term descriptions in this section include the following information:
a table of the error terms
description and significance of each error term
measurements affected by each error term
typical cause of failure for each error term
The same description applies to both the forward (F) and reverse (R) terms.
IMPORTANT
Data are listed here as a convenience only. Detailed instrument specifications
are listed in the embedded help in the network analyzer.
If Error Terms Seem Worse than Expected
To verify that the system still conforms to specifications, perform a system verification.
Refer to Chapter 3, “Tests and Adjustments”.
Table 8-1 Error Term Data
a
a. The data in this table are uncorrected system performance. The values apply over an environmental temperature
range of 25 °C ±5 °C, with less than 1 °C deviation from the calibration temperature
.
Parameter
Frequency Range
300 kHz
to
10 MHz
b
b. All values for 300 kHz to 10 MHz are typical.
10 MHz
to
1 GHz
1 GHz
to
3 GHz
3 GHz
to
5 GHz
5 GHz
to
11.5 GHz
11.5 GHz
to
13.5 GHz
Directivity –16 dB –28dB –25 dB –20dB –17dB –15dB
300 kHz
to
10 MHz
10 MHz
to
500 MHz
500 MHz
to
2 GHz
2 GHz
to
3 GHz
3 GHz
to
9 GHz
9 GHz
to
12.5 GHz
12.5 GHz
to
13.5 GHz
Source Match –18 dB –25dB –21dB –19dB –12dB –10dB –8dB
300 kHz
to
10 MHz
10 MHz
to
500 MHz
500 MHz
to
2 GHz
2 GHz
to
3 GHz
3 GHz
to
8 GHz
8 GHz
to
12.5 GHz
12.5 GHz
to
13.5 GHz
Load Match –17 dB –22dB –17dB –14dB –10dB –9dB –7dB
300 kHz
to
10 MHz
10 MHz
to
45 MHz
45 MHz
to
8 GHz
8 GHz
to
10.5 GHz
10.5 GHz
to
12.5 GHz
12.5 GHz
to
13.5 GHz
Crosstalk
c
c. All crosstalk values are typical. Measurement conditions: normalized to a thru, measured with two
shorts, 10 Hz IF bandwidth, averaging factor of 8, alternate mode, source power set to the lesser of
the maximum power out or the maximum receiver power.
–80dB –115dB –122dB –120dB –115dB –109dB

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5230A and is the answer not in the manual?

Agilent Technologies N5230A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5230A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals