3-34 Service Guide N5230-90014
Tests and Adjustments PNA Series Microwave Network Analyzers
Performance Tests (Agilent N7840A Software Package) N5230A
Dynamic Accuracy Test
Function of the Test: To measure the relative power linearity of the analyzers receivers.
Specification Tested: Test Port Input–Dynamic Accuracy
Equipment Used: The dynamic accuracy test set (Z5623A Option H01), a power meter
(E4418B or E4419B), power sensors, and two test cables.
Description of the Test:
1. The analyzers test ports are tested separately at a specific CW frequency and a
reference power level of 20 dBm.
2. The analyzer’s driving port is connected to the dynamic accuracy test set’s source port
and the analyzers receiving port is connected to the dynamic accuracy test set’s
receiver port. A power sensor is connected to the dynamic accuracy test set’s power
meter port.
3. The dynamic accuracy test set is used to input a signal from the driving port on the
analyzer. The input signal is routed through step attenuators to both the analyzer and
a power sensor.
4. The analyzer source and the dynamic accuracy test set create power levels that are
“deltas” from the reference power level of 20 dBm. Each delta is measured two ways:
by the power sensor and by the analyzers receiver under test.
5. The analyzer’s power level into the port under test is set to the 20 dBm reference
level.
6. The power level is measured with the power sensor: (P
mr
).
7. The power level is measured using the analyzers receiver under test: (P
pr
).
8. The power level into the analyzers port under test is stepped, in 5 dB steps, over the
range of 0 to 120 dBm. Each power level represents a “delta” from the reference power
level.
9. At each step, the power level is measured using the power meter: (P
md
).
10. At each step, the power level is measured using the receiver under test: (P
pd
).
11. The power error in dB = (P
mr
P
md
)(P
pr
P
pd
).
If the Analyzer Fails this Test:
If the analyzer fails this test, replace the A20 mixer brick and then repeat this test.
Refer to “Removing and Replacing the A20 Mixer Brick (QuintBrick)” on page 7-26.
If the analyzer still fails this test, replace the A5 SPAM board and repeat this test.
Refer to “Removing and Replacing the A5 through A10 Boards” on page 7-16.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5230A and is the answer not in the manual?

Agilent Technologies N5230A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5230A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals