Service Guide N5230-90014 3-35
PNA Series Microwave Network Analyzers Tests and Adjustments
N5230A Adjustments
Adjustments
These adjustments are firmware-driven tests that are used to fine-tune your analyzer.
If multiple adjustments are to be performed, perform them in the order listed.
“10 MHz Frequency Reference Adjustment” on page 3-35
“LO Power Adjustment” on page 3-36
“Source Calibration Adjustment” on page 3-37
“Receiver Calibration Adjustment” on page 3-38
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network
analyzers 10 MHz frequency reference on the A10 frequency reference board assembly.
Equipment Used for the Frequency Adjustment at 10 MHz
Procedure
NOTE
This adjustment typically adjusts to within ±0.01 ppm.
1. Connect the equipment as shown in Figure 3-13. Connect a GPIB cable between the
network analyzer and the frequency counter.
Figure 3-13 Setup for Adjustment of the 10 MHz Frequency Reference
2. On the System menu, point to Service, Adjustments, and then click
10 MHz Freq Adjust.
3. Ensure the GPIB settings are correct.
4. Click Begin Adj, and then follow the instructions as they are displayed.
Equipment Type Model or Part Number Alternate Model or Part Number
Cable, BNC, 50Ω, 24 inch 8120-1839 Any
Frequency counter 53151A, Option 001 Any that will measure a signal at 10 MHz.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5230A and is the answer not in the manual?

Agilent Technologies N5230A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5230A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals