Service Guide N5230-90014 5-15
PNA Series Microwave Network Analyzers Theory of Operation
N5230A Signal Separation Group Operation
Signal Separation Group Operation
The signal separation group divides the source incident signal into a reference path and a
test path. Refer to Figure 5-3 on page 5-15.
The reference signal are transmitted to the receiver group as the R1 and R2 inputs.
The test signal is transmitted through—and reflected from—the device under test
(DUT) and then is transmitted to the receiver group as the A and B inputs.
Control lines to this group are routed from the A16 test set motherboard.
In this section, the following assemblies are described:
A19 Switch/Splitter/Leveler/Amplifier/Multiplier (SSLAM)
A21 and A22 Test Port Couplers
A25 and A26 60-dB Source Step Attenuators (Option 1E1)
Front Panel Jumpers—Configurable Test Set (Option 014)
Figure 5-3 Passive Measurement Configuration