EasyManua.ls Logo

Agilent Technologies N5230A - Page 241

Agilent Technologies N5230A
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Guide N5230-90014 8-13
PNA Series Microwave Network Analyzers General Purpose Maintenance Procedures
N5230A Error Terms
Isolation (E
XF
and E
XR
)
Isolation, or crosstalk, is the uncorrected forward and reverse isolation error terms that
represent leakage between the test ports and the signal paths. The isolation error terms
are characterized by measuring transmission (S
21
, S
12
) with loads attached to both ports
during the measurement calibration.isolation errors affect transmission measurements
primarily where the measured signal level is very low.
The measurements most affected by isolation error terms are DUTs with large insertion
loss. Since these terms are low in magnitude, they are usually noisy (not very repeatable).
Typical Cause of Failure A loose cable connection or leakage between components in the
test set are the most likely cause of isolation problems.
After verifying the cable and its connections, suspect the analyzer switch splitter, step
attenuator, coupler, or receivers, and associated cabling.
To troubleshoot, refer to “Checking the Receiver Group” on page 4-45 or to “Checking the
Signal Separation Group” on page 4-42.

Table of Contents

Related product manuals