5Key Reference
122 N9342C/43C/44C User’s Guide
Sweep Time
Key access: [MEAS] > {SEM} > {Offset/Limits}
Specifies the sweep time for the currently selected
offset, and toggles this function between Auto and
Manual for each offset.
Res BW
Key access: [MEAS] > {SEM} > {Offset/Limits}
Specifies the resolution bandwidth used in
measuring the offset pair. When set to Auto, the
Res BW from the default tables for
“Offset/Limits" on page 119 are used. When set to
Man, the allowed range is 1 kHz to 1MHz.
Abs StartAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enables you to enter an absolute level limit at Start
Freq for the currently selected offset ranging from
–200.00 to +50.00 dBm with 0.01 dB resolution.
Abs StopAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enters an absolute level limit at Stop Freq for the
currently selected offset ranging from –200.00 to
+50.00 dBm with 0.01 dB resolution, and to toggle
this function between Couple and Man. If set to
Couple, this is coupled to Abs Start to make a flat
limit line. If set to Man, Abs Start and Abs Stop you
can enter different values to make a sloped limit
line.
Rel StartAmpt
Key access: [MEAS] > {SEM} > {Offset/Limits} > {More}
Enters a relative level limit at Start Freq ranging
from –150.00 to +50.00 dBc with 0.01 dB
resolution.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N9342C and is the answer not in the manual?

Agilent Technologies N9342C Specifications

General IconGeneral
BrandAgilent Technologies
ModelN9342C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals