TriScroll 600 Dry Scroll Vacuum Pump
D R A F T 4 / 2 9 / 0
4
19. Using the hemostat pliers, remove the large
spring, plunger guide, small spring and
check valve from the exhaust port in the
inboard housing.
The parts removed from the exhaust are:
Large spring
Plunger guide
Small spring
Check valve
TriScroll 600 Disassembly (continued)
NOTE The next step applies to all
models except the 610.
1
2
3
4