EasyManua.ls Logo

Allen-Bradley 1444-TSCX02-02RB - Page 135

Allen-Bradley 1444-TSCX02-02RB
550 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rockwell Automation Publication 1444-UM001D-EN-P - June 2018 135
Measurement Definition Chapter 4
Figure 56 - Time Constant Effect: True Peak Signal Detection, Signal Spike
A step change in a signal, as illustrated in Figure 57, can indicate a real change
in machine condition, such as from a loss of mass event. In such a case the
change could be significant, possibly cause damage, and can require immediate
action by operators or the protection system itself.
As with a signal spike, a True Peak measurement reacts almost instantaneously
to the stepped increase in magnitude. And within milliseconds after the
measurement, resolves to the same behavior of a steady state signal as described
previously.
Figure 57 - Time Constant Effect: True Peak-Peak Signal Detection, Step Change
RMS and Scaled Peak Measurements
The behavior of steady state measurements that apply RMS-based signal
detection is illustrated in Figure 58
and Figure 59. As illustrated, applying the
minimum time constant (0.1) can somewhat reduce measurement accuracy.
However, for RMS measurements, while a low TC does result in greater
variability in the measurement, the amount of the potential error, compared to
higher TC settings, is minimal.

Table of Contents