EasyManua.ls Logo

Allen-Bradley 825-P - Page 152

Allen-Bradley 825-P
266 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
152 Rockwell Automation Publication 825-UM004D-EN-P - November 2012
Chapter 11 Testing & Troubleshooting
2. Using the front panel or MPS Explorer software, record the Phase CT
Ratio and Phase Rotation setting values.
3. Set the phase current angles to apply balanced three-phase currents in
ac
cordance with the Phase Rotation setting. Refer to Figure48 on
page149.
4. S
et each phase current magnitude equal to the values listed in Column 1
of Ta
ble 85. Use the front panel to view the phase current values. The relay
displays the applied current magnitude times the CT Ratio setting.
For the minimum, mid-range, and maximum values, refer to the
I
e
setting range
for a given MCM converter module. These converter module current ranges are:
825-MCM2 (0.5…2.5 A)
825-MCM5 (1.0…5.0 A)
825-MCM20 (2.5…20 A)
825-MCM180 (20…180 A)
825-MCM420 (160…420 A)
825-MCM630N (160…630 A)
Current Unbalance Element Accuracy
1. Connect the current source to the MCM module as shown in Figure 48.
2. Using the front panel SET/SHOW function or MPS Explorer software,
rec
ord the Phase CT Ratio, Phase Rotation, and Motor FLA (
I
e
) setting
values.
3. Set the phase current angles to apply balanced three-phase currents in
ac
cordance with the Phase Rotation setting. See Figure 48.
4. Apply the appropriate magnitude for each phase current as shown in
colu
mn 1 of Tabl e 86
Table 85 - Phase Current Measuring Accuracy
|I|
Applied
(A secondary)
Expected
Reading
CTR1 x |I|
A–Phase
Reading
(A primary)
B–Phase
Reading
(A primary)
C–Phase
Reading
(A primary)
Minimum
Mid-range
Maximum
Table 86 - Current Unbalance Measuring Accuracy
|I| Applied (A secondary) Expected Reading (%) Actual Reading (%)
|IA| = 0.9 • FLA 7%
|IB| = FLA
|IC| = FLA
|IA| = 0.75 • FLA 17%
|IB| = FLA
|IC| = FLA

Table of Contents