EasyManuals Logo

Ametek 888 User Manual

Ametek 888
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
Analyzer Timing
| Appendix C-5
Period 13– Continuous Hold
All solenoid valves are de-energized during this period.
Period 14 – Continuous Sample
Process gas is continuously flowed through the sample cell during this
period.
Manual Operation
Set Modbus register 273 (Manual Operation) for the following values to
perform manual operation:
MANUAL IDLE = 0,
MANUAL ZERO = 1,
MANUAL CAL = 2,
MANUAL BLOWBACK = 3,
MANUAL MEASURE = 4,
MANUAL PERIOD1 = 10,
MANUAL PERIOD2 = 11,
MANUAL PERIOD3 = 12,
MANUAL PERIOD4 = 13,
MANUAL PERIOD5 = 14,
MANUAL PERIOD6 = 15,
MANUAL PERIOD7 = 16,
MANUAL PERIOD8 = 17,
MANUAL PERIOD9 = 18,
MANUAL PERIOD10 = 19,
MANUAL PERIOD11 = 20,
MANUAL PERIOD12 = 21,
MANUAL PERIOD13 = 22,
MANUAL PERIOD14 = 23

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 888 and is the answer not in the manual?

Ametek 888 Specifications

General IconGeneral
BrandAmetek
Model888
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals