Analyzer Timing
| Appendix C-5
Period 13– Continuous Hold
All solenoid valves are de-energized during this period.
Period 14 – Continuous Sample
Process gas is continuously flowed through the sample cell during this
period.
Manual Operation
Set Modbus register 273 (Manual Operation) for the following values to
perform manual operation:
MANUAL IDLE = 0,
MANUAL ZERO = 1,
MANUAL CAL = 2,
MANUAL BLOWBACK = 3,
MANUAL MEASURE = 4,
MANUAL PERIOD1 = 10,
MANUAL PERIOD2 = 11,
MANUAL PERIOD3 = 12,
MANUAL PERIOD4 = 13,
MANUAL PERIOD5 = 14,
MANUAL PERIOD6 = 15,
MANUAL PERIOD7 = 16,
MANUAL PERIOD8 = 17,
MANUAL PERIOD9 = 18,
MANUAL PERIOD10 = 19,
MANUAL PERIOD11 = 20,
MANUAL PERIOD12 = 21,
MANUAL PERIOD13 = 22,
MANUAL PERIOD14 = 23